The concurrent validity and reliability of the Leg Motion system for measuring ankle dorsiflexion range of motion in older adults
Romero Morales C, Calvo Lobo C, Rodríguez Sanz D, Sanz Corbalán I, Ruiz Ruiz B, López López D.. PeerJ, 2017-01-03
Topic: Brain imaging, connectivity and assessment
Dorsiflexion ROM limitation is associated with ankle injuries, foot pain, lower limb disorders, loss of balance, gait control disorders and fall risk in older adults. Results Mean values and standard deviations were as follows: Leg Motion (right 5.15 ± 3.08; left 5.19 ± 2.98), tape measure (right 5.12 ± 3.08; left 5.12 ± 2.80), goniometer (right 45.87° ± 4.98; left 44.50° ± 5.54) and inclinometer app (right 46.53° ± 4.79; left 45.27° ± 5.19). The paired t -test showed no significant differences between the limbs or between the test and re-test values. The test re-test reliability results for Leg Motion were as follows: the standard error of the measurement ranged from 0.29 to 0.43 cm, the minimal detectable difference ranged from 0.79 to 1.19 cm, and the intraclass correlation coefficients (ICC) values ranged from 0.97 to 0.98. Conclusions The results of the present study indicated that the Leg Motion device is a valid, reliable, accessible and portable tool as an alternative to the classic weight-bearing lunge test for measuring ankle dorsiflexion ROM in older adults.
Key findings
- Dorsiflexion ROM limitation is associated with ankle injuries, foot pain, lower limb disorders, loss of balance, gait control disorders and fall risk in older adults.
- Results Mean values and standard deviations were as follows: Leg Motion (right 5.15 ± 3.08; left 5.19 ± 2.98), tape measure (right 5.12 ± 3.08; left 5.12 ± 2.80), goniometer (right 45.87° ± 4.98; left 44.50° ± 5.54) and inclinometer app (right 46.53° ± 4.79; left 45.27° ± 5.19).
- The paired t -test showed no significant differences between the limbs or between the test and re-test values.
- The test re-test reliability results for Leg Motion were as follows: the standard error of the measurement ranged from 0.29 to 0.43 cm, the minimal detectable difference ranged from 0.79 to 1.19 cm, and the intraclass correlation coefficients (ICC) values ranged from 0.97 to 0.98.
- Conclusions The results of the present study indicated that the Leg Motion device is a valid, reliable, accessible and portable tool as an alternative to the classic weight-bearing lunge test for measuring ankle dorsiflexion ROM in older adults.
Study details
- Evidence strengthLow
How this study scored 41/100
Published in a peer-reviewed journal. Scored zero for study design, sample size, effect reported, studied in people and randomised / blinded because the record does not state it. We have no data on journal's citation record, times cited, citations per year, impact for its field and year, cited by studies here, still being cited, senior author's track record, institution's track record and multi-centre collaboration, so those were left out: it earned 24 of the 58 points we could judge it on, which is 41 out of 100.
| Component | Points | Why |
|---|---|---|
| Journal tier | 7 / 10 | Tier 3 — peer-reviewed journal (PeerJ) |
| Journal's citation record | — / 4 | No journal record from OpenAlex — left out of this score |
| Study design | 0 / 12 | No study design recorded — scores zero, not guessed (not stated in the record — scored 0) |
| Strength of evidence | 1 / 5 | Rated “low” in our summary |
| Peer-review validation | 10 / 10 | Peer-reviewed. Published as an article in a named peer-reviewed journal. |
| How recent | 2 / 5 | 2017 — within 10 year(s) of the newest study here |
| Sample size | 0 / 4 | Participant count not stated — scores zero (not stated in the record — scored 0) |
| Effect reported | 0 / 3 | No effect size reported — scores zero (not stated in the record — scored 0) |
| Studied in people | 0 / 3 | Not stated whether this was studied in people — scores zero (not stated in the record — scored 0) |
| Randomised / blinded | 0 / 2 | No randomisation or blinding stated — scores zero (not stated in the record — scored 0) |
| Times cited | — / 7 | No citation count (not found in OpenAlex) — left out of this score |
| Citations per year | — / 5 | No citation count or no year — left out of this score |
| Impact for its field and year | — / 6 | No field-weighted citation impact (OpenAlex has none yet) — left out of this score |
| Cited by studies here | — / 6 | Not yet checked against this library's citation graph — left out of this score |
| Still being cited | — / 4 | No yearly citation counts — left out of this score |
| Senior author's track record | — / 4 | Senior author's record not available — left out of this score |
| Institution's track record | — / 3 | No institution listed in OpenAlex — left out of this score |
| Multi-centre collaboration | — / 3 | Institutions not listed — left out of this score |
| Indexed in PubMed | 2 / 2 | Indexed in PubMed (PMID 28070457) |
| Free to read | 2 / 2 | Free to read |
| Points | 24 / 58 | Scored on 11 of 20 parts. We have no data for the rest, so they are left out: 24 of the 58 points we could judge it on is 41 out of 100. |
| Total | 41 / 100 | — |
What each of these 20 components measures, and what this score is not →
Read the original
Free copy unknown: no free-copy check has an answer for this record yet. That does not mean it is paywalled.
Cite this study APA, Vancouver, BibTeX · download .bib or .ris
Built from the record alzhub holds for this paper, not looked up. Volume, issue and page numbers are not held in the record. Check the citation against the original before you use it.
APA 7th
Romero Morales, C., Calvo Lobo, C., Rodríguez Sanz, D., Sanz Corbalán, I., Ruiz Ruiz, B., & López López, D. (2017). The concurrent validity and reliability of the Leg Motion system for measuring ankle dorsiflexion range of motion in older adults. PeerJ. https://doi.org/10.7717/peerj.2820
Vancouver
Romero Morales C, Calvo Lobo C, Rodríguez Sanz D, Sanz Corbalán I, Ruiz Ruiz B, López López D. The concurrent validity and reliability of the Leg Motion system for measuring ankle dorsiflexion range of motion in older adults. PeerJ. 2017. doi:10.7717/peerj.2820. PMID: 28070457.
BibTeX
@article{romeromorales2017-28070457,
author = {Romero Morales, C. and Calvo Lobo, C. and Rodríguez Sanz, D. and Sanz Corbalán, I. and Ruiz Ruiz, B. and López López, D.},
title = {{The concurrent validity and reliability of the Leg Motion system for measuring ankle dorsiflexion range of motion in older adults}},
journal = {PeerJ},
year = {2017},
doi = {10.7717/peerj.2820},
pmid = {28070457},
url = {https://doi.org/10.7717/peerj.2820}
}